The most accurate optical single-ion clock worldwide

Scientists from the Physikalisch-Technische Bundesanstalt (PTB) reduce the measurement uncertainty of their ytterbium clock down to 3 ∙ 10–18

Categories:
  • Press releases
  • Science
  • Press
  • Optical atomic clocks
  • SI-Basiseinheit (für Kapitel "Forschung & Entwicklung" – bei Bedarf eine Kategorie auswählen)
Press release
10.02.2016

Schematic representation: Measuring the influence of thermal ambient radiation on the frequency of the trapped ion: the "clock laser" (blue beam) excites the trapped ion (yellow) with a special pulse sequence. The resonance frequency of the ion is influenced by infrared radiation (here by an infrared laser, red beam). This can be measured by means of the clock laser. (Fig.: PTB)

The definition and realization of the SI unit of time, the second, is currently based on cesium atomic clocks. Their "pendulum" consists of atoms which are excited into resonance by microwave radiation (1010 Hz). It is regarded as certain that a future redefinition of the SI second will be based on an optical atomic clock. These have a considerably higher excitation frequency (1014 to 1015 Hz), which makes them much more stable and more accurate than cesium clocks.

The accuracy now achieved with the ytterbium clock is approximately a hundred times better than that of the best cesium clocks. To develop their clock, the researchers from PTB exploited particular physical properties of Yb+. This ion has two reference transitions which can be used for an optical clock. One of these transitions is based on the excitation into the so-called "F state" which, due to its extremely long natural lifetime (approx. 6 years), provides exceptionally narrow resonance. In addition, due to the particular electronic structure of the F state, the shifts of the resonance frequency caused by electric and magnetic fields are exceptionally small. The other reference transition (into the D3/2 state) exhibits higher frequency shifts and is therefore used as a sensitive "sensor" to optimize and control the operating conditions. Another advantage is that the wavelengths of the lasers required to prepare and excite Yb+ are in a range in which reliable and affordable semiconductor lasers can be used.

The decisive factor for the last leap in accuracy was the combination of two measures: firstly, a special procedure was conceived for the excitation of the reference transition. With this procedure, the "light shift" of the resonance frequency caused by the exciting laser is measured separately. This information is then used to immunize the excitation of the reference transition against the light shift and its possible variation. Secondly, the frequency shift caused by the thermal infrared radiation of the environment (which is relatively small for the F state of Yb+ anyway) was determined with a measurement uncertainty of only 3 %. For this purpose, the frequency shift caused by laser light and its intensity distribution at the ion's location were measured at four different wavelengths in the infrared range.

Another particular property of the F state of Yb+ is the strong dependence of the state energy on the value of the fine-structure constant (the elementary fundamental constant of electromagnetic interac-tion) and on the anisotropy effects in the interaction between electrons and certain potential forms of the so-called dark matter which plays an important part in the present cosmologic standard model. Comparisons between Yb+ clocks and other highly accurate optical clocks are currently probably the most promising way of verifying theories from this area of "new physics" in the lab.
es/ptb


Contact

Dr. Christian Tamm, Senior Scientist, Department 4.4, Subject Area "Optical frequency standards", phone: +49 (0)531 592–4415, e-mail: Opens window for sending emailchristian.tamm(at)ptb.de

 

Scientific publication

N. Huntemann, C. Sanner, B. Lipphardt, Chr. Tamm, E. Peik: Opens external link in new windowSingle ion atomic clock with 3 ∙ 10–18 uncertainty. Physical Review Letters 116, 063001 (2016) 