Der "magnetisch stillste Raum der Welt" steht auf dem Gelände der PTB in Berlin.

The "magnetically best shielded room on earth" is the size of an apartment block and is located on the site of the Physikalisch-Technische Bundesanstalt (PTB), Institute Berlin. Magnetic fields such as that of the earth are kept out here as effectively as nowhere else. Such ideal conditions allow the tiny magnetic fields of, e.g., the human heart to be measured. This was the motivation for the American National Institute of Standards and Technology (NIST) to ask PTB to jointly test a newly developed optical magnetic field sensor. It is based on a physical principle very different from SQUIDs, which are usually applied for biomagnetic field measurements. The optical sensor does not need advanced cooling and is the size of a lump of sugar. A high-quality measurement of the human heart signal was demonstrated using this optical sensor. The sensor's suitability was thus proven for biomagnetic measurements in the picotesla range. In future magnetocardiographic measurement devices - to be used as a supplement or an alternative to the ECG - could become simpler and less expensive.

Up until now you had to cool as much as you could for biomagnetic measurements! This was necessary as SQUIDs, superconducting quantum interference devices, work optimally at -269 degrees Celsius and can only then fulfil their purpose of measuring tiny magnetic fields. SQUIDs are the best suited sensors to record the magnetic fields arising during the electrical activity of the human heart. A magnetocardiogram (MCG) can be compiled supplementing a conventional electrocardiogram (ECG). (The same applies to the magnetoencephalogram, MEG, which is a recording of the magnetic field of the brain.) Yet to use SQUIDs requires well-shielded rooms and complicated cooling systems. The latter might become obsolete in the future if the optical magnetometer developed by NIST continues to fulfil expectations.

The optical sensor is a byproduct of the development of miniaturized atomic clocks, one of NIST's current key research areas. As biomagnetic research is not established at NIST, the scientists turned to their colleagues at PTB in Berlin. PTB is one of the few metrological state institutes in the world with a large-scale biomedical research program. The combination of highly advanced equipment such as the magnetically shielded room (BMSR-2) with the experience in testing new sensors on human subjects made PTB the ideal partner for NIST.

Using the optical magnetometer, PTB experts measured the magnetic field of the human heart and relaxation curves of magnetic nanoparticles. Both are important routine laboratory measurements normally applying SQUIDs. For a direct comparison of the results, SQUID reference data were recorded simultaneously with the optical magnetometer using the multichannel SQUID device installed in the BMSR-2.

The impressive quality of the data obtained confirms the suitability of these optical micromagnetometers in the picotesla range. This is the range for magnetic field measurements of the heart, but further improvements are needed to measure the even tinier fields of the human brain. Compared to SQUIDs, the optical sensors exhibited - as expected - a significantly higher noise level. The distinct advantage of these sensors lies, however, in their small design (< 1 cm3). This allows them to be installed at a small distance from the magnetic source in order to increase the signal strength. The room temperature operation of the sensor and its small size will allow a much simpler system design compared to other sensor types suitable for this measurement range. The fabrication of the sensor by use of microsystem technology could enable simple and cost-effective mass production. NIST plans the development of a multichannel system on the basis of the existing technology. Further tests of these sensors at PTB are planned.

Contact:
Dr. Tilman Sander-Thömmes, PTB Working Group 8.21 Biomagnetism,
Tel.: +49 30 3481-7436,
e-mail: tilmann.sander-thoemmes(at)ptb.de

Contact

Head of Press and Information Office

Dr. Dr. Jens Simon