Department 6.3 | Working Group 6.32 |
Working Group 6.31 | Working Group 6.34 |
The following services are covered by the International Committee for Weights and Measures (CIPM) Mutual Recognition Arrangement (see eleven CMC entries).
All irradiations are carried out in compliance with the relevant ISO standards, ISO 29661 and ISO 6980 for beta radiation, ISO 4037 and ISO/TS 18090-1 for photon radiation and ISO 20956 for photon radiation at Low Dose Rates.
Contacts
Beta radiation:
Dr. Rolf Behrens
phone: +49 (0) 531-592-6340
CaliSievertBeta(at)ptb.de
Photon radiation:
Dr. Hayo Zutz
phone: +49 (0) 531-592-6310
CaliSievertPhoton(at)ptb.de
Photon radiation at Low Dose Rates:
Dr. Faton Krasniqi
phone: +49 (0) 531-592-6320
Measuring instruments
- Without readout device:
for example ionization chambers - With readout device:
for example scintillation detectors, counting tubes or semiconductor detectors - Passive dosemeters
Measured quantities
- H'(0,07), Hp(0,07), H'(3) and Hp(3), for beta and photon radiation
- H*(10) and Hp(10) for photon radiation
- Ka for photon radiation at Low Dose Rates
Phantoms for Personal dosemeter irradiations
- ISO PMMA rod phantom for the measured quantity Hp(0,07)
- ISO water cylinder phantom for the measured quantity Hp(3)
- ISO water slab phantom for the measured Hp(10)
Forms for ordering calibration, testing or irradiation
Beta radiation:
Order form for the irradiation of passive dosemeter badges
Order form for calibration or test of active dosemeters or secondary standards
Photon radiation:
Order form for the irradiation of passive dosemeter badges
Order form for calibration or test of active dosemeters or secondary standards
Photon radiation at Low Dose Rates:
- For ordering, please contact directly (see above)
Radiation qualities
Beta radiation:
- Beta radiation fields in accordance with ISO 6980‑1 from the following isotopes:
Pm‑147, Kr‑85, Sr‑90/Y‑90, Ru‑106/Rh‑106
Photon radiation:
- X-radiation: ISO 4037‑1 N-, W- and H‑series from 8 keV to 250 keV
- Gamma radiation: ISO 4037‑1 S‑Co (Co‑60) and S‑Cs (Cs‑137)
- High energy photon radiation: ISO 4037‑1 R‑C (4.4 MeV) and R-F (6 MeV to 7 MeV)
Photon radiation of Low Dose Rates:
- Gamma radiation: Cs‑137, Co‑60, Co‑57, Am‑241, Ra‑226
- Zero effect of dosimetry systems
- Secondary cosmic radiation
Dose ranges and dose rate ranges
Beta radiation:
- Dose range: 1 μSv to 1 Sv
- Dose rate range: 30 μSv/h to 0.2 Sv/h
Photon radiation:
(depending on irradiation facility):
- Dose range: 0.1 μSv to 100 Sv
- Dose rate range: 0.1 μSv/h to 10 Sv/h
Photon radiation of Low Dose Rates:
- Dose range: 0.1 μSv to 2 μSv
- Dose rate range: 0.01 μSv/h to 0.4 μSv/h
Costs