Logo of the Physikalisch-Technische Bundesanstalt

Free-air ionization chamber (cylindrical ionization chamber)

Technical data:

Electrode diameter:40 cm
High voltage:3 kV
Tube voltage range:30 kV - 300 kV
Reference volumes:2.5 cm3 -180 cm3
Leakage current:< 50 fA
Relative expanded uncertainty:0.64% (k = 2)