PTB and NIST performed a comparison for high accuracy continuous wave optical power measurements in the kilowatt regime. The NIST carried out measurements with a power meter relying on photon momentum, while PTB performed the measurements with a modified off-the-shelf thermal power meter. The non-absorbing photon momentum measurement approach permits the two power meters to measure the same laser beam optical path simultaneously, resulting in a direct comparison of the meters supported by an optical system to accommodate differences in instrument settling times. The results show agreement within the expanded uncertainties for each instrument. NIST and PTB illustrate a degree of equivalence of 0.49% with an expanded uncertainty of 1.37% (k = 2) for an average result across all power levels.
Authors:
Kyle Rogers1, Paul Williams1, Marcel Pastuschek2, Holger Lecher2, Stefan Kück2,
Marco López2 and John Lehman1
1 Sources and Detectors Group, US National Institute of Standards and Technology, Boulder, CO, United States of America
2 Optics Division, Physikalisch-Technische Bundesanstalt, Braunschweig, Germany
For further information:
Kyle Rogers et al, Metrologia 61 025006, 2024. https://doi.org/10.1088/1681-7575/ad2538