Direct Comparison of Josephson Waveforms using an AC Quantum Voltmeter |
| Author(s): |
R. Behr, L. Palafox, G. Ramm, H. Moser and J. Melcher |
| Journal: |
IEEE Trans. Instrum. Meas. IM-56 |
| Year: |
2007 |
| Volume: |
Nr. 2 |
| Pages: |
S. 235-238 |
| Abstract: |
A new method for calibrating ac voltage waveforms based on measuring the difference relative to a synchronously synthesized waveform supplied from a programmable Josephson array is presented. In order to test the accuracy of this approach, the differences between two Josephson waveforms have been measured at the 1.2-V level for frequencies up to 5 kHz. The uncertainties achieved are at least a factor of ten lower than those from the direct sampling method. Below 400 Hz, this new method gives an uncertainty of 5 10-8 (k = 1), while at 1 kHz, the uncertainty is about 2 107 (k = 1). |